Tube 06SA
1. PERFORMANCE
1) Measuring range :20-1,000 ppm 5-50 ppm
Number of pump strokes 1(100ml) 4(400ml)
2) Sampling time :3 minutes/1 pump stroke
3) Detectable limit :2 ppm(400mr)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Dark brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Pottasium disulphate palladate(II)is reduced, and Palladium is liberated.
CO+K2Pd(SO3)2→K2(SO3)2PdCO
K2(SO3)2PdCO→CO2+SO2+Pd+K2SO3
4. CALIBRATION OF THE TUBE
STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
20
GAS To pump
50
100
200
300
400
500
600
700
800
900
1000
ppm
co
??????
?? ?? ??
106SA
CARBON MONOXIDE
(NOTE)
When the concentration is below 50 ppm, 4 pump strokes can be used to detemine the lower concentration.
Following formula is available for the actual concentration.
Actual concentration=1/4×Temperature corrected value
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
1,000 870 930 1,000 1,030 1,060
900 780 840 900 930 960
800 690 750 800 830 850
700 610 680 700 720 720
600 520 560 600 620 640
500 430 470 500 520 540
400 350 370 400 410 430
300 260 280 300 310 320
200 180 190 200 210 220
100 90 100 100 100 110
Substance ppm Interference ppm Coexistence
Ethylene 5,000 Pale grey stain is produced. 5,000 The top of discoloured layer becomes
unclear and higher readings are given.
Whole layer is discoloured to Greyish
Hydrogen 5,000 Greyish yellow stain is produced. 5,000 yellow and the top of discoloured layer
becomes unclear.
Acetylene 1.5 Dark green stain is produced. C×Oc1o/n5c. Higher readings are given.
Sulphur dioxide 100 Original colour is faded. 〃〃
Nitrogen dioxide Tahffee catcecdu.racy of readings is not 〃〃
www.r-biopharm.com.cn
www.r-biopharm.com.cn
請(qǐng)輸入賬號(hào)
請(qǐng)輸入密碼
請(qǐng)輸驗(yàn)證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購(gòu)買(mǎi)風(fēng)險(xiǎn),建議您在購(gòu)買(mǎi)產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。