The new Model 212 Scrubberless Ozone Monitor uses the proven ultraviolet (UV) absorption method with an innovative gas-phase scrubber technology to measure ozone virtually free from interferences. In the reference mode, the instrument uses the reaction between ozone and nitric oxide (NO) to selectively remove ozone from the sample. This reaction is orders of magnitude faster than with any other ambient gas, resulting in the removal of only ozone and not potentially interfering compounds that absorb UV light. The instrument is ideally suited for measurements of ozone in heavily polluted air where interference is likely from particulates, mercury, or VOCs. The ozone monitor combines the stability of our proven dual beam ozone monitors with our selective scrubber technology to provide accurate ozone measurements in the most polluted air.
The gas-phase scrubber technology uses NO to quantitatively react with ozone and remove it from the sample. This reaction has long been used as a gas phase titration for the measurement of either NO or O3. In the Scrubberless Ozone Monitor, a low concentration of NO is added to the sample to generate ozone-free, reference air. The light intensity is measured in the reference and sample modes, and the concentration of ozone is calculated directly from Beer's Law. Nitric oxide can be supplied from an external NO cylinder and will provide over a year of gas supply. Alternatively, nitric oxide can be generated inside the instrument using an optional photolytic NO generator and an external cylinder of nitrous oxide (N2O), which is non-toxic, non-corrosive and relatively inexpensive.
All 2B Tech ambient air monitors offer remote instrument operation via RS-232, choice of averaging time (2 s to 1 hr), analog output (0-2.5 V), current output (4-20 mA), internal data logger, and optional flash memory. Our data display and graphing software allows continuous display of data which can be saved for offline analysis
Features
- Virtually interference-free measurement of ambient ozone
- Free of signal quenching interferences inherent to chemiluminescence methods
- DewLine™ for elimination of water vapor interference, a unique feature of 2B Tech instruments
- Inexpensive, simple, and robust instrument design in a 19” rack mount enclosure
- Touchscreen display
- Internal data logger and USB flash memory
- Dual beam optical bench with high precision
- Stable calibration (calibration suggested annually)
- Electronic sample and nitric oxide flow control
- Internal long-life sample pump
- Under evaluation as a Federal Equivalent Method (FEM)
- Convenient user interface, including remote operation via RS-232 communication
- Optional internal NO generator using non-toxic, noncorrosive nitrous oxide (N2O)
Options
- GPS for continuous logging of latitude, longitude and altitude
- External particle filter
Specifications
Measurement Principle | UV Absorption at 254 nm; Dual Beam |
Linear Dynamic Range | 0.5 ppb to 2.0 ppm |
Resolution | 0.1 ppb |
Precision (1σ; rms noise) | Greater of 0.5 ppb or 1% of reading for 10-s average |
Accuracy | Greater of 1.0 ppb or 2% of reading |
Limit of Detection (2σ) | 1.0 ppb |
NIST-Traceable Calibration | Yes |
Flow Rate (nominal) | ~2 Liter/min |
Flow Rate Requirement | >1.2 L/min |
Baseline Drift | <1 ppb/day |
Sensitivity Drift | <0.5%/day |
Measurement Time, Frequency | 2 s, 0.5 Hz |
Response Time, 100% of Step Change | 4 s for 2-s averaging; 20 s for 10-s averaging |
Averaging Times | 2 s, 10 s, 1 min, 5 min, 1 hr |
Internal Data Logger Capacity | 14,336 lines (10 s avg. = 1.4 days; 1 min avg = 10 days; 5 min avg = 1.4 mo; 1 hr avg = 1.6 yr); Virtually unlimited capacity with USB Flash Drive |
Ozone Units | ppb, ppm, pphm, µg m-3, mg m-3 |
Pressure Units | mbar, torr |
Temperature Units | °C, K |
T and P Corrected | Yes |
Operating Temperature Range | 10 to 50 °C |
Operating Altitude Range | ~0-30 km (~30-1000 mbar) |
Power Requirement; Supplied by Battery or 110/220 VAC Power Pack | 12 V dc or 120/240 V ac, 1.1 A at 12 V, 13 watt; 8 watt with Touch Screen off |
Size | Rackmount: 17" w x 14.5" d x 5.5" h (43 x 37 x 14 cm) |
Weight | 14.3 lb (6.5 kg) |
Analog Inputs for Internal Logging of Other Instruments | 3 Analog Inputs, 0-2.5 V; For example could log external T, P and RH |
Data Outputs | Touch Screen Display, USB, RS232, 0-2.5 V Analog, Optional 4-20 mA Analog |
Data Transfer Baud Rates | 1200, 2400, 4800, 19200 |
Output Ranges | User Defined Scaling Factor in Menu |
DewLine™ | Yes |
Long Life Pump | Yes, 20,000 hr |
Flow Meter | Yes |
Options | GPS, Lamp Heater; Bluetooth for wireless data transmission |
System Includes
- Model 211 Scrubberless Ozone Monitor
- AC Power Adapter (100-240 VAC to 12 VDC) with Country-Specific Plug
- Zeroing Cartridge
- Operation Manual on CD
- Calibration Data and NIST-Traceable Calibration Certificate
- Instrument Birth Certificate
- One Year Warranty
相關產(chǎn)品
- UV-106-W水中溶解臭氧分析儀
- 美國Eco Sensors OS-6美國Eco Sensors OS-6臭氧控制器
- 美國ECO Sensor OS-4美國ECO Sensor OS-4臭氧開關監(jiān)測器
- 美國ECO SensorEZ-1X臭氧監(jiān)測變送器
- 美國ECO SensorC-30ZX臭氧監(jiān)測器
- 美國ECO SensorA-22美國ECO SensorA-22臭氧檢測儀
- Q45H/64水中溶解臭氧分析儀
- UV-106美國ECO UV-106系列紫外臭氧檢測儀
- Model Mini-Hicon高濃度臭氧分析儀Model Mini-Hicon高濃度臭氧分析儀
- IN-2000-L2-LCIN-2000-L2-LC臭氧分析儀
- UV-100美國ECO UV-100紫外分光光度法臭氧檢測儀
- POM便攜式紫外臭氧檢測儀POM便攜式紫外分光光度法臭氧檢測儀
- Model 211免擦洗紫外臭氧分析儀Model 211免擦洗紫外臭氧分析儀
- Model 205雙光束紫外臭氧分析儀Model 205雙光束紫外臭氧分析儀
- Model 202紫外臭氧分析儀Model 202紫外臭氧分析儀
- Model 106美國2B Model 106紫外法臭氧檢測儀
- A-21ZX袖珍型臭氧檢測儀A-21ZX袖珍型臭氧檢測儀
請輸入產(chǎn)品關鍵字:
郵編:
聯(lián)系人:王經(jīng)理
電話:86-21-67159386
傳真:
手機:18621801483
留言:發(fā)送留言
個性化:www.shwi.com.cn
網(wǎng)址:www.lab17.com.cn
商鋪:http://www.syzwkj.com/st235541/