Megger PAT320 Technical Specifications
Electrical Supply Range | |
230V AC | 230V AC ±10% |
110V AC | 110V AC ±10% |
Supply Measurement | |
Voltage | ±2% ± 1V |
Frequency | ±1% ± 0.1Hz |
Bond Test | |
Open Circuit Voltage | 9V AC ±10% ±0.1V (Supply: 230V 50Hz) |
10A Bond Test Current | 10A rms ±20% ±0. into 0.1Ω |
2 Bond Test Current | 2 rms ± 5% ± 0. into 0.1Ω |
Earth Bond Resistance Accuracy | ±5% ± 3 digits (0 to 0.49Ω) ±5% ± 5 digits (0.5 to .99Ω) |
Earth Bond Resistance Resolution | 10mΩ (0 to 1.99Ω) |
Display Range | 0 to 1.99Ω |
Bond Test Nulling | Up to 1.00Ω |
Adjustable Test Duration | User selectable from 1 sec to 20 sec |
Continuity Test | |
Note | The continuity test is a DC test performed automatically in both positive and negative directions. The average of the two results is shown |
Continuity Test Voltage | 4.0V DC -0% +10% (open circuit) |
Continuity Test Current | 200mA -0% + 10% ± 5mA (into 2Ω load) |
Continuity Resistance Accuracy | ±5% ±3 digits (0 to 0.49Ω) ±5% ±5 digits (0.5 to .99Ω) |
Continuity Resistance Resolution | 10mΩ (1 to 19.99Ω) |
Display Range | 0 to 19.99Ω |
Continuity Test Nulling | Up to 9.99Ω |
Test Duration | User selectable from 1 sec to 20 sec |
Insulation Test | |
Insulation Test Voltage | 250V DC -0%/+25% open circuit 500V DC -0%/+25% open circuit (500V DC across 0.5MΩ) |
Insulation Resistance Accuracy (230V) | ±2% ±5 digits (0 to 19.99MΩ) ±5% ± 10 digits (20 to 99.99MΩ) |
Insulation Resistance Accuracy (110V) | ±5 ±5 digits (0 to 10MΩ) |
Insulation Resistance Resolution | 0.01MΩ (0.10 to 99.99MΩ) |
Display Range | 0 to 99.99MΩ |
Test Duration | User selectable from 1 sec to 1 minutes |
Substitute Leakage Test | |
Test Voltage | 40V AC ±10% |
Test Frequency | Nominal mains 50/60Hz |
Leakage Current Accuracy | ±5% ± 5 digits |
Leakage Current Resolution | 0.01mA |
Display Range | 0 to 19.99mA |
Test Duration | User selectable from 1 sec to 1 minute Reading corrected to 230V + 10% or 110V +10% AC |
Differential Leakage Current | |
Test Voltage | Nominal mains 110V/230V AC |
Test Frequency | Nominal mains 50/60Hz |
Differential Leakage Current Accuracy | ±5% ± 5 digits |
Differential Leakage Current Resolution | 0.01mA |
Display Range | 0 to 19.99mA Reading corrected to 230V +10 % or 110V + 10% AC |
Touch Current Test | |
Test Voltage | Nominal mains 1150/230V AC |
Test Frequency | Nominal mains 50/60Hz |
Touch Current Accuracy | ±5% ±5 digits |
Touch Current Resolution | 0.01mA |
Display Range | 0 to 10mA |
Test Duration | User selectable from 1 sec to 5 sec Reading corrected to 230V + 10% or 110V + 10% AC |
Operational Test | |
Test Voltage | Nominal mains 110V/230V AC |
Accuracy | ±5% ±10 digits (0 to 99VA) ±5% ±50 digits (100 to 999VA) ±5% ±100 digits (1000 to 3700VA) |
Resolution | 1VA (0 to 3700VA) |
Display Range | 0 to 3.99KVA Reading corrected to 230V or 110V AC Results show load VA |
Extension Lead Test | |
Polarity Test Voltage | 12V |
Polarity | Leak OK Live neutral S/C Live neutral reversed Live/neutral O/C |
Portable RCD Test | |
RCD Test Voltage | Nominal mains 110/230V |
RCD Test Frequency | 50Hz |
Test Current Accuracy | -8% to -2% (1/2 x I) ±2% to +8^ (1 x 1,5 x I) |
Trip Time Accuracy | ±1% ±5 digits |
Trip Time Resolution | 0.1ms |
Display Range | 0 to 1999ms (1/2 x I) 0 to 300ms (1 x I) 0 to 40ms (5 x I) |
Fuse Test | |
Test Voltage | 3.3V |
Warning | Audible Beep if fuse is ok |