性猛交XXXX乱大交派对,四虎影视WWW在线观看免费 ,137最大但人文艺术摄影,联系附近成熟妇女

| 注冊(cè)| 產(chǎn)品展廳| 收藏該商鋪

行業(yè)產(chǎn)品

當(dāng)前位置:
上海島韓實(shí)業(yè)有限公司>>技術(shù)文章>>KERN 572/573/KB/DS/FKB/FCB/KBJ

KERN 572/573/KB/DS/FKB/FCB/KBJ

閱讀:1457        發(fā)布時(shí)間:2015-5-6

KERN & Sohn GmbH
KERN 572/573/KB/DS/FKB/FCB/KBJ
Version 7.1
03/2014
GB
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471
GB
KERN 572/573/KB/DS/FKB/FCB/KBJ
Version 7.1 03/2014
Operating instruction
Precision/compact balances and platform scales
Table of contents :
1 MODE - MENUE ....................................................................................................... 3
2 Technical data ......................................................................................................... 5
2.1 KERN 572 ..............................................................................................................................................5
2.2 KERN 573 ..............................................................................................................................................8
2.3 KERN KB/KBJ ......................................................................................................................................9
2.4 KERN DS .............................................................................................................................................13
2.5 KERN FCB ..........................................................................................................................................18
2.6 KERN FKB ..........................................................................................................................................19
3 Fundamental information (general) ..................................................................... 22
3.1 Intended use .......................................................................................................................................22
3.2 Inappropriate use ...............................................................................................................................22
3.3 Guarantee ...........................................................................................................................................22
3.4 Monitoring the test substances .......................................................................................................22
4 Fundamental safety information .......................................................................... 23
4.1 Observe the information in the operating instructions .................................................................23
4.2 Staff training .......................................................................................................................................23
5 Transport and storage .......................................................................................... 23
5.1 Acceptance check ..............................................................................................................................23
5.2 Packaging ...........................................................................................................................................23
6 Unpacking, installation and commissioning ...................................................... 24
6.1 Place of installation, place of use ....................................................................................................24
6.2 Unpacking ..........................................................................................................................................24
6.2.1 Installation ............................................................................................................................................24
6.3 Mains supply ......................................................................................................................................24
6.4 Battery operation FKB .......................................................................................................................25
6.5 Connecting peripheral equipment ...................................................................................................25
6.6 Initial start-up .....................................................................................................................................25
6.7 Adjustment .........................................................................................................................................25
6.8 Adjusting (go to 7.2.6) .......................................................................................................................26
6.8.1 Adjusting for verification (KERN 573; FKB-M; KB-NM; FCB-M; DS-M) ...............................................26
6.9 Verification .........................................................................................................................................27
7 Operation ............................................................................................................... 28
7.1 Display control panel ........................................................................................................................28
7.2 Operation ............................................................................................................................................29
7.3 Display background illumination .....................................................................................................35
7.4 Interface RS 232 C .............................................................................................................................35
7.5 RS 232 C Data output via interface RS 232 C .................................................................................36
7.5.1 Description of the data transfer............................................................................................................37
7.5.2 Numerator ............................................................................................................................................37
7.6 Printer .................................................................................................................................................37
7.7 Underfloor weighing ..........................................................................................................................38
8 Maintenance, upkeep, disposal ........................................................................... 39
8.1 Cleaning ..............................................................................................................................................39
8.2 Maintenance, upkeep ........................................................................................................................39
8.3 Disposal ..............................................................................................................................................39
9 Troubleshooting .................................................................................................... 40
10 Declaration of conformity ..................................................................................... 41
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  2
1  MODE - MENUE
Models 572 / KB / DS / FKB / FCB non-verifiable:
Important!
The modified settings - as well as the adjustment - must be stored when switching-off, via the
ON/OFF button.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  3
Models 573-NM; KB-NM; DS-M; FKB-M; KBJ-NM:
Important!
The modified settings - as well as the adjustment - must be stored when switching-off, via the
ON/OFF button.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  4
2  Technical data
2.1 KERN 572
KERN  572-30  572-31  572-32  572-33
Readability (d)  0.001 g  0.001 g  0.001 g  0.01 g
Weighing range (max)  240 g  300 g  420 g  1600 g
Taring range (subtractive)  240 g  300 g  420 g  1600 g
Reproducibility  0.001 g  0.002 g  0.002 g  0.01g
Linearity  ± 0.003 g  ± 0.005 g  ± 0.005 g  ± 0.03 g
Smallest piece weight  0.001 g  0.001 g  0.001 g  0.01 g
Adjustment points  50/100/
200/240 g
50/100/
200/300 g
100/200/
300/400 g
0.5/1.0/
1.5/1.6 kg
Recommended adjusting
weight F1 (not supplied)
200 g  200 g +100 g  200 g + 200 g  1 kg + 500 g
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  4 hours  2 hours
Housing (B x D x H) mm  180 x 310 x 90
Vibration filter  yes
Weighing plate stainless
steel mm
? 106 ? 106  ? 106  ? 150
Units  see menu
Weight kg (net)  2.3  2.3  2.3  2.3
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  5
KERN  572-35  572-37  572-39  572-43
Readability (d)  0.01 g  0.01 g  0.01 g  0.1 g
Weighing range (max)  2400 g  3000 g  4200 g  10000 g
Taring range (subtractive)  2400 g  3000 g  4200 g  10000 g
Reproducibility  0.01 g  0.02 g  0.02 g  0.1g
Linearity  ± 0.03 g  ± 0.05 g  ± 0.05 g  ± 0.3 g
Smallest piece weight  0.01 g  0.01 g  0.01 g  0.1 g
Adjustment points  0.5/1.0/
2.0/2.4 kg
1.0/1.5/
2.0/3.0 kg
1.0/2.0/
3.0/4.0 kg
2/5/10 kg
Recommended adjusting
weight F1 (not supplied)
2 kg  2 kg + 1 kg  2 kg + 2 kg  10 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  4 hours  2 hours
Housing (B x D x H) mm  180 x 310 x 90
Vibration filter  yes
Weighing plate stainless
steel mm
? 150  ? 150  ? 150  160 x 200
Units  see menu
Weight kg (net)  2.3  2.3  2.3  2.7
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  6
KERN  572-45  572-49  572-55  572-57
Readability (d)  0.05 g  0.1 g  0.05 g  0.1 g
Weighing range (max)  12000 g  16000 g  20000 g  24000 g
Taring range (subtractive)  12000 g  16000 g  20000 g  24000 g
Reproducibility  0.05 g  0.1 g  0.1 g  0.1g
Linearity  ± 0.15 g  ± 0.3 g  ± 0.25 g  ± 0.3 g
Smallest piece weight  0.05 g  0.1 g  0.05 g  0.1 g
Adjustment points  2/5/10/12 kg  5/10/15/16 kg  5/10/15/20 kg  5/10/15/20/24 kg
Recommended adjusting
weight F1 (not supplied)
10 kg  10 kg + 5 kg  20 kg  20 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours  2 hours
Housing (B x D x H) mm  180 x 310 x 90
Vibration filter  yes
Weighing plate stainless
steel mm
160 x 200  160 x 200  160 x 200  160 x 200
Units  see menu
Weight kg (net)  2.7  2.7  2.7  2.7
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  7
2.2  KERN 573
KERN  573-34NM  573-46NM
Accuracy class  II  II
Readability (d)  0.01 g  0.1 g
Verification value (e)  0.1 g  1 g
Weighing range (max)  650 g  6500 g
Minimum load (Min)  0.5 g  5 g
Taring range (subtractive)  650 g  6500 g
Reproducibility  0.01 g  0.1 g
Linearity  ± 0.03 g  ± 0.3 g
Smallest piece weight  0.01 g  0.1 g
Adjustment points  200g/500g/600g  2/5/6 kg
Recommended adjusting
weight F1 (not supplied)
500 g + 100 g  5 kg + 1 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 30 °C
Warm-up time  2 hours  2 hours
Housing (B x D x H) mm  180 x 310 x 90
Vibration filter  yes
Weighing plate stainless
steel mm
? 150  160 x 200
Units  g, kg  g, kg
Weight kg (net)  2.3  2.7
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  8
2.3  KERN KB/KBJ
KERN  KB 120-3N  KB 240-3N  KB 360-3N
Readability (d)  0.001 g  0.001 g  0.001 g
Weighing range (max)  120 g  240 g  360 g
Taring range (subtractive)  120 g  240 g  360 g
Reproducibility  0.001 g  0.001 g  0.002 g
Linearity  ± 0.003 g  ± 0.003 g  ± 0.005 g
Smallest piece weight  0.001 g  0.001 g  0.001 g
Adjustment points  20/50/100/120 g  100/150/
200/240 g
100/200/
300/360 g
Recommended adjusting
weight F1 (not supplied)
100 g  200 g  200 g + 100 g
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  4 hours
Housing (B x D x H) mm  167 x 250 x 85
Vibration filter  yes
Weighing plate stainless
steel mm
? 81  ? 81  ? 81
Units  see menu
Weight kg (net)  1  1  1
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  9
KERN  KB 1200-2N  KB 2000-2N  KB 2400-2N  KB 3600-2N
Readability (d)  0.01 g  0.01 g  0.01 g  0.01 g
Weighing range (max)  1 200 g  2 000 g  2 400 g  3 600 g
Taring range (subtractive)  1 200 g  2 000 g  2 400 g  3 600 g
Reproducibility  0.01 g  0.01 g  0.01 g  0.02 g
Linearity  ± 0.03 g  ± 0.03 g  ± 0.03 g  ± 0.05 g
Smallest piece weight  0.01 g  0.01 g  0.01 g  0.01 g
Adjustment points  0.2/0.5/1.0/1.2 kg 0.5/1.0/1.5/2.0 kg 1.0/1.5/2.0/2.4 kg 1.0/2.0/3.0/3.6 kg
Recommended adjusting
weight F1 (not supplied)
1 000 g  2 000 g  2 000 g  2 kg + 1 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours  4 hours
Housing (B x D x H) mm  167 x 250 x 85
Vibration filter  yes
Weighing plate stainless
steel mm
130 x 130  130 x 130  130 x 130  130 x 130
Units  see menu
Weight kg (net)  1.5  1.5  1.5  1.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  10
KERN
KB
10000-1N
KB
10k0.05N
Readability (d)  0.1 g  0.05 g
Weighing range (max)  10 000 g  10 000 g
Taring range (subtractive)
10 000 g  10 000 g
Reproducibility
0.1 g  0.05 g
Linearity
0.3 g  0.15 g
Smallest piece weight
0.1 g  0.05 g
Adjustment points
2/5/10 kg  2/5/10 kg
Recommended adjusting
weight F1 (not supplied)
10 kg  10 kg
Stabilization time (typical)
3 sec.  3 sec.
Permitted environmental tem-
perature
+10 °C ... + 40 °C  +10 °C ... + 40 °C
Warm-up time
2 hours  2 hours
Housing (B x D x H) mm
167 x 250 x 85  167 x 250 x 85
Weighing plate stainless steel
mm
150 x 170  150 x 170
Units
see menu  see menu
Vibration filter
yes  yes
Weight kg (net)
1.7  1.7
Data interface
yes (RS232)  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  11
KERN
KB
650-2NM
KB
6500-1NM
KBJ
650-2NM
Accuracy class  II  II  II
Readability (d)  0.01 g  0.1 g  0.01 g
Verification value (e)  0.1 g  1 g  0.1 g
Weighing range (max)  650 g  6 500 g  650 g
Minimum load (Min)  0.5 g  5 g  0.5 g
Taring range (subtractive)  650 g  6 500 g  650 g
Reproducibility  0.01 g  0.1 g  0.01 g
Linearity  ± 0.03 g  ± 0.3 g  ±0.03 g
Smallest piece weight  0.01 g  0.1 g  0.01 g
Adjustment points  200/500/600 g  2/5/6 kg  -
Recommended adjusting
weight F1 (not supplied)
600 g  6.0 kg Int. adjustment
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 30 °C
Warm-up time  2 hours  2 hours  2 hours
Housing (B x D x H) mm  167 x 250 x 85
Vibration filter  yes
Weighing plate stainless
steel mm
130 x 130  150 x 170  130 x 130
Units  g, kg, ct
Weight kg (net)  1.5  1.7  2.1
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  12
2.4  KERN DS
KERN
DS
3K0.01S
DS
5K0.05S
DS
8K0.05
DS
10K0.1S
Readability (d)  0.01 g  0.05 g  0.05 g  0,1 g
Weighing range (max)  3 000 g  5 000 g  8 000 g  10.000 g
Taring range (subtractive)  3 000 g  5 000 g  8 000 g  10.000 g
Reproducibility  0.02 g  0.05 g  0.05 g  0.1 g
Linearity  ± 0.05 g  ± 0.15 g  ± 0.15 g  ±0.3 g
Smallest piece weight  0.01 g  0.05 g  0.05 g  0.1 g
Adjustment points  1/2/3 kg  2/5/10 kg  2/4/5/7/8 kg  2/5/10 kg
Recommended adjusting
weight F1 (not supplied)
3 kg  10 kg  5 kg + 2 kg  10 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours  2 hours
Housing (B x D x H) mm  228 x 228 x 70  228 x 228 x 70  315 x 305 x 70  228 x 228 x 70
Vibration filter 
Weighing plate stainless
steel mm
228 x 228  228 x 228  315 x 305  228 x 228
Units  see menu  see menu  see menu  see menu
Weight kg (net)  5.5  5.5  7.5  5.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  13
KERN
DS
16K0.1
DS
20K0.1
DS
30K0.1
Readability (d)  0.1 g  0.1 g  0.1 g
Weighing range (max)  16 000 g 20 000 g  30 000 g
Taring range (subtractive)  16 000 g  20 000 g  30 000 g
Reproducibility  0.1 g  0.1 g  0.2 g
Linearity  ± 0.3 g  ± 0.3 g  ± 0.5 g
Smallest piece weight  0.1 g  0.1 g  0.1 g
Adjustment points  5/10/15/16 kg  5/10/15/20 kg  10/15/20/30 kg
Recommended adjusting
weight F1 (not supplied)
10 kg + 5 kg  20kg  20 kg + 10 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours
Housing (B x D x H) mm  315 x 305 x 70
Vibration filter  yes
Weighing plate stainless
steel mm
315 x 305  315 x 305  315 x 305
Units  see menu  see menu  see menu
Weight kg (net)  7.5  7.5  7.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  14
KERN
DS
36K0.2
DS
30K0.1L
DS
36K0.2L
DS
60K0,2
Readability (d)  0.2 g  0.1 g  0.2 g  0.2 g
Weighing range (max)  36 000 g  30 000 g  36 000 g  60 000 g
Taring range (subtractive)  36 000 g  30 000 g  36 000 g  60 000 g
Reproducibility  0.2 g  0.2 g  0.2 g  0.4 g
Linearity  ± 0.6 g  ± 0.5 g  ± 0.6 g  ± 1.0 g
Smallest piece weight  0.2 g  0.1 g  0.2 g  0.2 g
Adjustment points  10/15/20/
30/36 kg
10/15/20/
30 kg
10/15/20/
30/36 kg
20/30/50/60 kg
Recommended adjusting
weight F1 (not supplied)
20 kg + 10 kg  20 kg + 10 kg  20 kg + 10 kg  50 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours  2 hours
Housing (B x D x H) mm  315 x 305 x 70  450 x 350 x 115
Vibration filter  yes
Weighing plate stainless
steel mm
315 x 305  450 x 350
Units  see menu  see menu  see menu  see menu
Weight kg (net)  7.5  9.5  9.5  9.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  15
KERN
DS
65K0.5
DS
100K0.5
DS
150K1
Readability (d)  0.5 g  0.5 g  1 g
Weighing range (max)  65 000 g  100 000 g 150 000 g
Taring range (subtractive)  65 000 g  100 000 g  150 000 g
Reproducibility  0.0 g  0.5 g  1 g
Linearity  ± 1.5 g  ± 1.5 g  ± 3 g
Smallest piece weight  0.5 g  0.5 g  1 g
Adjustment points  20/30/50/60 kg  20/50/100 kg  50/100/150 kg
Recommended adjusting
weight F1 (not supplied)
50 kg  50 kg + 50 kg
3 x 50 kg
3 x 50 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours
Housing (B x D x H) mm  450 x 350 x 115
Vibration filter  yes
Weighing plate stainless
steel mm
450 x 350
Units  see menu  see menu  see menu
Weight kg (net)  9.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  16
KERN
DS
65K1M
Accuracy class  II
Readability (d)  1 g
Verification value (e)  10 g
Weighing range (max)  65 000 g
Minimum load (Min)  50 g
Taring range (subtractive)  65 000 g
Reproducibility  1 g
Linearity  ± 2 g
Smallest piece weight  1 g
Adjustment points  20/30/50/
60 kg
Recommended adjusting
weight F1 (not supplied)
50 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 30 °C
Warm-up time  2 hours
Housing (B x D x H) mm  450 x 350 x 115
Vibration filter 
Weighing plate stainless
steel mm
450 x 350
Units  see menu
Weight kg (net)  9.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  17
2.5  KERN FCB
KERN
FCB
6K0.02B
FCB
12K0.05B
FCB
12K0.1B
FCB
24K0.1B
FCB
24K0.2B
Readability (d)  0.02 g  0.05 g  0.1 g  0.1 g  0.2 g
Weighing range (max)  6 000 g  12 000 g  12 000 g  24 000 g  24 000 g
Taring range (subtrac-
tive)
6 000 g  12 000 g  12 000 g  24 000 g  24 000 g
Reproducibility  0.04 g  0.05 g  0.1 g  0.1 g  0.2 g
Linearity ±  0.1 g  0.15 g  0.3 g  0.3 g  0.6 g
Smallest piece weight  0.02 g  0.05 g  0.1 g  0.1 g  0.2 g
Adjustment points  2/5/6 kg  2/5/10/12 kg  2/5/10/12 kg
5/10/15/
20/24 kg
5/10/15/
20/24 kg
Recom. adjustment
weight F1 
(not added)
5 kg  10 kg 10 kg  20 kg 20 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typi-
cal)
3 sec.
Permitted environmen-
tal temperature
+ 10 °C ... + 40 °C
Warm-up time  2 hours  2 hours  2 hours  2 hours  2 hours
Vibration filter  yes
Housing 
(B x D x H) mm
270 x 345 x 106 270 x 345 x 106 270 x 345 x 106 270 x 345 x 106 270 x 345 x 106
Weighing plate, stain-
less steel
253x228  253x228  253x228  253x228  253x228
Units  see menu
Vibration filter  yes
Weight kg (net)  3.3  3.3  3.3  3.3  3.3
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  18
2.6  KERN FKB
KERN
FKB
6K0.02
FKB
8K0.05
FKB
16K0.05
FKB
16K0.1
Readability (d)  0.02 g  0.05 g  0.05 g  0.1 g
Weighing range (max)  6 000 g  8 000 g 16 000 g  16 000 g
Taring range (subtractive)  6 000 g  8 000 g  16 000 g  16 000 g
Reproducibility  0.04 g  0.05 g  0.1 g  0.1g
Linearity  ± 0.1 g  ± 0.15 g  ± 0.25 g  ± 0.3 g
Smallest piece weight  0.02 g  0.05 g  0.05 g  0.1 g
Adjustment points  2/4/5/6 kg  2/4/5/7/8 kg  5/10/15/16 kg  5/10/15/16 kg
Recommended adjusting
weight F1 (not supplied)
5 kg  5 kg + 2 kg  10 kg + 5 kg  10 kg + 5 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time
2 hours  2 hours  2 hours  2 hours
Housing (B x D x H) mm  350 x 390 x 120
Vibration filter  yes
Weighing plate stainless
steel mm
340 x 240  340 x 240  340 x 240  340 x 240
Units  see menu  see menu  see menu  see menu
Weight kg (net)  6.5  6.5  6.5  6.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  19
KERN
FKB
36K0.1
FKB
36K0.2
FKB
65K0.2
FKB
65K0.5
Readability (d)  0.1 g  0.2 g  0.2 g  0.5 g
Weighing range (max)  36 000 g  36 000 g  65 000 g  65 000 g
Taring range (subtractive)  36 000 g  36 000 g  65 000 g  65 000 g
Reproducibility  0.2 g  0.2 g  0.4 g  0. g
Linearity  ± 0.5 g  ± 0.6 g  ± 1.0 g  ± 1.5 g
Smallest piece weight  0.1 g  0.2 g  0.2 g  0.5 g
Adjustment points  10/20/30/36 kg  10/20/30/36 kg  20/30/50/60 kg  20/30/50/60 kg
Recommended adjusting
weight F1 (not supplied)
20 kg + 10 kg  20 kg + 10 kg  50kg  50kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 40 °C
Warm-up time  4 hours  2 hours  4 hours  2 hours
Housing (B x D x H) mm  350 x 390 x 120
Vibration filter  yes
Weighing plate stainless
steel mm
340 x 240  340 x 240  340 x 240  340 x 240
Units  see menu  see menu  see menu  see menu
Weight kg (net)  6.5  6.5  6.5  6.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  20
KERN
FKB
8K0.1M
FKB
65K1M
Accuracy class  II  II
Readability (d)  0.1 g  1 g
Verification value (e)  1.0 g  10 g
Weighing range (max)  8 000 g  65 000 g
Minimum load (Min)  5.0 g  50 g
Taring range (subtractive)  8 000 g  65 000 g
Reproducibility  0.1 g  1 g
Linearity  ± 0.3 g  ± 3 g
Smallest piece weight  0.1 g  1 g
Adjustment points  2/5/7/8 kg  20/30/50/60 kg
Recommended adjusting
weight F1 (not supplied)
5 kg + 2 kg  50 kg
Humidity of air  max. 80% rel. (non-condensing)
Stabilization time (typical)  3 sec.
Permitted environmental
temperature
+10 °C ... + 30 °C  +10 °C ... + 30 °C
Warm-up time  2 hours  2 hours
Housing (B x D x H) mm  350 x 390 x 120
Vibration filter  yes
Weighing plate stainless
steel mm
340 x 240  340 x 240
Units  see menu  see menu
Weight kg (net)  6.5  6.5
Data interface  yes (RS232)
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  21
3  Fundamental information (general)
It is essential to read through and observe the complete operating instructions before in-
stalling and commissioning.
3.1  Intended use
The balance you have acquired serves to determine the weighing value of the material to
be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be
weighed is manually and carefully placed in the centre of the weighing plate. The weighing
value can be read off after a stable weighing value has been obtained.
3.2  Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are re-
moved or added to the material to be weighed, incorrect weighing results can be displayed
due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a
container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the measuring
equipment.
Be sure to avoid impact shock and overloading the balance in excess of the prescribed
maximum load rating (max.), minus any possible tare weight that is already present. This
could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not explosion-
proof.
Structural alterations may not be made to the balance. This can lead to incorrect weighing
results, faults concerning safety regulations as well as to destruction of the balance.
The balance may only be used in compliance with the described guidelines. Varying areas
of application/planned use must be approved by KERN in writing.
3.3  Guarantee
The guarantee is not valid following 
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
3.4  Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight must
be checked at regular intervals within the scope of quality assurance. For this purpose, the
answerable user must define a suitable interval as well as the nature and scope of this
check. Information is available on KERN’s home page (www.kern-sohn。。com) with regard
to the monitoring of balance test substances and the test weights required for this. Test
weights and balances can be adjusted quickly and at a reasonable price in KERN’s ac-
credited DKD calibration laboratory (return to national normal).
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  22
4  Fundamental safety information
4.1  Observe the information in the operating instructions
Carefully read this operation manual before setup and com-
missioning, even if you are already familiar with KERN balanc-
es.
Versions in other languages are non-binding translations.
The only binding version is the original document in German.
4.2  Staff training
The device may only be operated and looked after by trained members of staff.
5  Transport and storage
5.1  Acceptance check
Please check the packaging immediay upon delivery and the device during unpacking
for any visible signs of external damage.
In the case of visible damages please obtain a signature from the bearer to serve as con-
firmation of damage. Do not make alterations to the goods and packaging, do not remove
any consignment parts. Report the damage to the parcel service immediay in writing (
within 24 hours).
5.2  Packaging
Please retain all parts of the original packaging in case it should be necessary to return
items at any time.
Only the original packaging should be used for return consignments.
Before despatch, disconnect all attached cables and loose/movable parts, remove weigh-
ing plate.
Apply any intended transport security devices. Secure all parts, e.g. glass windshield,
weighing plate, power unit etc., to prevent slipping and damage.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  23
6  Unpacking, installation and commissioning
6.1  Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be achieved
under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and pre-
cisely.
Therefore please observe the following at the place of installation:
-  Place the balance on a firm, level surface;
-  Avoid extreme heat as well as temperature fluctuation caused by installing next to a
radiator or in the direct sunlight;
-  Protect the balance against direct draughts due to open windows and doors;
-  Avoid jarring during weighing;
-  Protect the balance against high humidity, vapours and dust;
-  Do not expose the device to extreme dampness for longer periods of time. Inad-
missible bedewing (condensation of air moisture on the device) can occur if a cold
device is taken into a significantly warmer environment. In this case, please accli-
matise the device for approx. 2 hours at room temperature after it has been discon-
nected from the mains.
-  Avoid static charging of the material to be weighed, weighing container and wind-
shield.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields
occur as well as due to static charging and instable power supply. It is then necessary to
change the location.
6.2  Unpacking
Carefully remove the balance from its packaging, remove the plastic wrapping and posi-
tion the balance in its intended working location.
6.2.1 Installation
Install the balance in such a fashion that the weighing plate is absoluy horizontally.
6.3  Mains supply
Electric power supply is by means of the external mains supply circuit. The printed voltage
level must comply with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to ap-
proval by Kern.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  24
6.4 Battery operation FKB
? To insert the batteries (6 x 1.5 V) remove the battery compartment cover. Remove it
with the help of a coin.
? In the each battery tube insert three batteries in the same polarity sense.
? Screw down again the battery cover.
To save the battery, the background illumination can be switched off (see chap. 7.3).
Moreover the AUTO-OFF function can be activated (see chap. 7.2.10).
If the battery voltage drops below a critical value for operational safety, this will be indicat-
ed in the display with the “BATT LOW“ information.
6.5  Connecting peripheral equipment
The balance must be disconnected from the mains before connecting or disconnecting
additional equipment (printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These have
been ideally coordinated to your balance.
6.6  Initial start-up
A warm-up time of 2 hours stabilises the measured values after switching on.
The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on adjusting.
6.7  Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated – in compliance with the underlying physical weighing prin-
ciple - to the existing acceleration due to gravity at its place of location ( only if the balance
has not already been adjusted to the location in the factory). This adjustment process
must be carried out during the initial start-up, after change in location and variation of sur-
rounding temperature. It is also recommendable to adjust the balance periodically during
weighing operation in order to obtain exact measured values.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  25
6.8  Adjusting (go to 7.2.6)
Using a precision weight, the accuracy of the balance can be checked at any time and
adjusted.
Warning:
With verified balances, this precision adjustment is not possible.
Exception: model KBJ with internal adjustment weight.
Adjustment procedure:
Check that the surrounding conditions are stable.
A short warm-up time of about 15 minutes is recommended for stabilisation.
6.8.1 Adjusting for verification (KERN 573; FKB-M; KB-NM; FCB-M; DS-M)
General Information:
Prior to the procedure of verification the balance has to be adjusted.
Remark: The adjusting is only possible when not being blocked by the adjusting switch.
The switch for this step is located at the bottom of the housing between the two
turning feet.
At the model FKB the cover plate from the housing bottom must be removed.
Switch to the right Adjusting function is released.
This setting has to be selected before starting the
adjusting.
Executing the adjusting according to chapter 7.2
“Operation“. Model KBJ see 6.8.2
After the adjusting has been completed successful-
ly it is necessary to turn the adjusting switch to the
left for blocking.
Switch to the left  Adjusting function is blocked.
After the adjusting has been completed successful-
ly the adjusting has to be blocked by switching to
the left.
The balance is now prepared for the procedure of
verification.
After the verification the adjusting switch has to be
secured by a verification mark against access.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  26
6.8.2 Adjustment of the balance KBJ with internal adjustment weight
After expiring of the warm-up time (wait > 299s), the balance KBJ is automatically
adjusted, whenever switched-on again. Additionally, at any time another adjustment can
be released manually via the keyboard.
In order to carry out an exact adjustment, the environmental influences (draught/ vibration)
must be kept to an absolute minimum. The weighing pan must be placed without load. An
adjustment weight is not necessary here.
Sequence of the manual adjustment:
?  Press the Mode key:
?  Display: Cal? acknowledge by Yes button
?  Display: int. Cal Weight / Adjustment motor starts up
?  Display: Auto-Calibration
?  Display: Calibration done
?  Display: 0.00 g
6.9 Verification
General introduction:
According to EU directive 90/384/EEC balances must be verified if they are used as fol-
lows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the medi-
cal and pharmaceutical laboratory.
c) For official purposes.
d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions
An EU type approval exists for balances described in their technical data as verifiable. If a
balance is used where obligation to verify exists as described above, it must verified and
re-verified in regular intervals.
Re-verification of a balance is carried out according to the respective national regulations.
The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Verification of the balance is invalid without the "seal".
Balances with obligation to verify must be taken out of operation if:
? ?  The weighing result of the balance is outside the error limit. Therefore, in reg-
ular intervals load balance with known test weight (ca. 1/3 of the max. load) and
compare with displayed value.
? ?  The reverification deadline has been exceeded.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  27
7  Operation
7.1 Display control panel
Keyboard
ON / OFF
Print weighing result
in MODE: No / descending
in % and counting mode:
Form reference
in MODE: Yes / ascending
MODE configuration
(see mode structure diagram page 2)
Formula weighing
in MODE: to left
Change g-pieces
Unit changeover
in MODE: to right
Tare
in MODE: back in weighing operation
Display-Symbol  Meaning
==OVERLOAD==
Overload: Weighing range has been exceeded
= ============
Underload: Weighing range has been fallen short of
<< .
In counting and % mode: part too light
? .
Automatic tare active / In verification operation zero display
PTA .
Preselect tare Tare pre-selection active
D .
Difference in % during percentage weighing
Net .
Net weight of the components during formula preparation
SUM .
Gross weight of several components during formula preparation
? .
Balance is in counting mode and currently displaying the weight value of
the counting amount
.
For multirange balance in verification operation indication of range
Wait > 299 s Warm-up time if verifiable (switch-on drift monitoring)
III
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  28
7.2 Operation
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  29
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  30
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  31
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  32
7.2.13 Animal weighing:
? Activate via mode/animal weigh. Y/N (4xMode)
? Put container without measured object on the weighing surface and tare
? Place the measured object (animal) in the container and start with REF/YES key
? The value averaging in shown in the display with AVG as well as by extinguishing of
the character blocks and the subsequently fixed display value.
? Printing and deleting of the average value can be triggered by the PRINT key
? Simple deleting is possible by pressing the TARE key.
?
7.2.14 Swap:
? Strong filtering
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  33
7.2.15 Contents of the form printout:
MODE / PRINTER / PRINTERFORM
Var 1 Type xx = Gross  MODE
Var 1 Row xx XXxx
MODE
Var 1 Type 02 = Gross
Var 1 Row 02
Var 1 Col 14
Gross value active
Line 2
Column 14
REF / YES  MODE
Var 2 Type xx = Net
Var 3 Type xx = Tare
Var 4 Type xx = Pretare
Var 5 Type xx = Pcs / %
Var 6 Type xx = * / Mw
Var 7 Type xx = Numerator
Printerform
Var 1 Col xx
Printout “Gross“ line 2 / column 14
xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx 1. line
xxxxxxxxxxxxxG 15000.0g xxxxxx 2. line
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  34
7.3 Display background illumination
After having switched on your balance and the zero display, press the ?MODE“ button to
select the menu point ?Backlight“. Acknowledge by pressing the ?YES“ button in order to
switch on the permanent background illumination. By pressing the ?NO“ button, the back-
ground illumination is switched off.
If the background illumination shall be switched off time-controlled (to save the battery),
press the ?MODE“ button to select the menu point ?Backlight auto“ and confirm by press-
ing the ?YES“ button. The background illumination will be switched off automatically 10 sec
after having reached a stable weighing value.
7.4 Interface RS 232 C
Technical Data
? 8-bit ASCII Code
? 1 start bit, 8 data bits, 1 stop bits, no parity bit
? Baud rate adjustable to 2400, 4800 and 9600 baud (default), 19200 baud
? Sup-D- 9 pol. is necessary .
? When working with an interface correct operation is secured only if the corresponding
KERN-interface-cable (max. 2m) is used.
Description of the jack
Sup-D- 9 pol.
5  1
9 6
Pin 2:  Transmit data
Pin 3:  Receive data
Pin 5:  Signal ground
Baudrate
The Baud rate for the data transfer is adjusted with the MODE-key. The following example
demonstrates how to set the Baud rate 4800.
Select Baud rate  Display
1. Press MODE-key repeatedly until "PRINTER" is displayed.
2. Press YES-key.
3. Press MODE-key repeatedly until the desired Baud rate
appears (for instance 4800 Baud).
4. Press YES-key to select 4800 Baud. The tick-mark (X) con-
firms the new setting.
5. Press MODE-key repeatedly until the balance displays in
grams again, or press tare key.
PRINTER?
2400 Baud
4800 Baud
4800 Baud X
0.0 g
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  35
7.5 RS 232 C Data output via interface RS 232 C
RS 232 C Data output via interface RS 232 C
General information
As a condition for the data transfer between the balance and a peripheral device (for in-
stance printer, PC ...) both devise have to be set on the same interface parameter (for in-
stance baud rate, parity ...).
There are 5 methods for the data output via RS 232 C
Data output via PRINT-Key
The printing process can be released by the PRINT-key. In this case the settings AUTO-
PRINT and AUTPRINT PC should be deselected.
AUTOPRINT (Data output, after having loaded the balance)
The setting AUTOPRINT is in the PRINTER-routine, and there it can be selected or dese-
lected. When AUTOPRINT is active the actual weighing value will be sent via the RS 232
interface when the balance has been unloaded and then loaded after having achieved the
stability.
AUTOPRINT PC (Continuous data output)
The setting AUTPRINT PC is in the PRINTER-routine, and there it can be selected or de-
selected. When AUTOPRINT PC is active the actual weighing values will be sent continu-
ously via the RS 232 interface.
Data output by transfer of remote controls
The following functions can be released by the remote controls that will be transferred as
ASCII signs to the balance.
t  Tare.
w  a weighing value (or unstable) is sent via RS 232 interface.
s  a stable weighing value is sent via RS 232 interface.
If the balance receives the command w or s, it acts without printing delay.
Output on bar code printer
The data transfer mode has to be set on ?Barcode“.
As bar code printer a Zebra printer model LP2824 is provided.
Take into account that the output format of the balance is fixedly defined and cannot be
changed.
The printer format is stored in the printer, i.e. in case of a failure the printer cannot be
changed with a new one from factory, previously it is necessary that KERN installs the re-
spective software.
The Zebra printer and the balance must be connected to the delivered interface cable
when they are switched off.
After switching-on both appliances, and after reaching the status ready-for-operation, a
label will be printed out when pressing the  key.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  36
7.5.1 Description of the data transfer
Structure of each data transfer:
Bit-Nr.  1  2  3  4  5  6  7  8  9  10  11  12  13  14  15  16  17  18
B*  N  N  N  B  B  B  B  B  0
.
0  0  E  E  E  CR  LF
N  =  Numerator
B*:  =  Blank or in the range of zero point.
B, 0,  . , g: =  Blank or weighing value with unit, depending on the load on the weighing plate.
E  =  Unit
CR:  =  Carriage Return
LF:  =  Line Feed
7.5.2 Numerator
The Numerator is found under the menu point “Printer" and can be activated and deac-
tivated. When editing data with the use of the print option, the range of the numerator will
increase by one place.
7.6 Printer
With the serial data output RS 232 a printer can be connected.
The printout shows the weight in grams. When the counting mode is selected the number
of pieces or the weight is printed. When the percent mode is selected, the percentage or
the weight will be printed. Press The PRINT-key to print weighing results. Select the enu-
merator to number the weighing continuously. Turn off the balance or use the CLEAR
function to Reset the enumerator to (000).
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  37
7.7 Underfloor weighing
Objects which, because of their size or shape, cannot be put on the scale, can be weighed
by means of underfloor weighing.
Proceed as follows:
?  Switch off the balance.
?  Turn the balance over, without loading the balance plate.
?  Open the cover plate on the base of the balance.
?  Hang on the hook for underfloor weighing .
?  Place the balance over an opening.
?  Hang the item to be weighed on the hook and carry out weighing.
! CAUTION !
Take care that the hooks used for the underfloor weighing are stable enough to hold
the goods which you wish to weigh (risk of breakage).
Always make sure that there are no living beings or materials below the load that
could be injured or damaged.
! NOTE !
After completing the underfloor weighing, the opening in the floor of the balance
must be closed again (dust protection).
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  38
8  Maintenance, upkeep, disposal
8.1 Cleaning
Please disconnect the device from the operating voltage before cleaning.
Only use a cloth dampened with mild suds and not aggressive cleaning agents (solvents
or similar). Please ensure that fluids are not able to get into the device and rub off using a
clean, soft cloth.
Loose sample residue/powder can be removed carefully using a brush or hand vacuum
cleaner.
Remove any spilt material to be weighed immediay.
8.2 Maintenance, upkeep
The device may only be opened by trained service engineers authorised by KERN.
Disconnect from the mains supply before opening.
8.3 Disposal
The operating company shall dispose of the packaging and the device in compliance with
the valid national or regional law of the operating location.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  39
9  Troubleshooting
The balance should be switched off for a short time following an interruption in the pro-
gramme sequence and disconnected from the mains supply. It is then necessary to repeat
the weighing process from the beginning.
Help:
Interruption  Possible cause
Weight display is not illuminated.  ? The balance is not switched on.
? The mains supply connection has been inter-
rupted (mains cable not plugged in/faulty).
? Power supply interrupted. .
The weight display changes continu-
ally
? Draught/air movement
? Table/floor vibrations
? The weighing plate is in contact with foreign
matter.
? Electromagnetic fields / static charging
(choose different location/switch off interfering
device if possible)
The weighing result is obviously in-
correct
? The balance display is not set to zero
? Adjustment is no longer correct.
? Great fluctuations in temperature.
? Electromagnetic fields / static charging
(choose different location/switch off interfering
device if possible)
Switch the balance off if other error messages should appear and then switch on again.
Contact the manufacturer if the error message does not disappear.
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  40
10  Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
: info@kern-sohn.de
0049-[0]7433- 9933-0
0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Declaration of conformity
EC-Konformit?tserkl?rung  EC-Declaration of -Conformity
EC- Déclaration de conformité  EC-Declaración de Conformidad
EC-Dichiarazione di conformità  EC-Conformiteitverklaring
EC- Declara??o de conformidade  EC- Prohlá?ení o shode
EC-Deklaracja zgodno?ci  EC-Заявление о соответствии
D
Konformit?ts-
erkl?rung
Wir erkl?ren hiermit, dass das Produkt, auf das sich diese Erkl?rung bezieht,
mit den nachstehenden Normen übereinstimmt.
GB
Declaration of
conformity
We hereby declare that the product to which this declaration refers conforms
with the following standards.
CZ
Prohlá?ení o
shode
Tímto prohla?ujeme, ?e vyrobek, kterého se toto prohlá?ení tyká, je v souladu
s ní?e uvedenymi normami.
E
Declaración de
conformidad
Manifestamos en la presente que el producto al que se refiere esta
declaración está de acuerdo con las normas siguientes
F
Déclaration de
conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la
présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di
conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferi-
sce è conforme alle norme di seguito citate.
NL
Conformiteit-
verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking
heeft, met de hierna vermelde normen overeenstemt.
P
Declara??o de
conformidade
Declaramos por meio da presente que o produto no qual se refere esta
declara??o, corresponde às normas seguintes.
PL
Deklaracja
zgodno?ci
Niniejszym o?wiadczamy, ?e produkt, którego niniejsze o?wiadczenie dotyczy,
jest zgodny z poni?szymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
Electronic Balance:  KERN 572 / 573 / FCB / FKB / KB..N / DS
Mark applied  EU Directive  Standards
2004/108/EC
EN 61000-6-3:2007
EN 55011:2009/A1:2010
EN 55022:2010/AC:2011
EN 55024:2010
EN 45501:1992-10+AC:1993-08
OIML R 76-1:2006
2006/95/EC  EN 60950
Signatur
Signature
Datum
Date
08.04.2013
Ort der Ausslung
Place of issue
72336 Balingen  Albert Sauter
KERN & Sohn GmbH
Gesch?ftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, . +49-[0]7433/9933-0
+49-[0]7433/9933-149, : info@kern-sohn.com, Internet: www.kern-sohn。。com
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  41
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
: info@kern-sohn.de
0049-[0]7433- 9933-0
0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Declaration of conformity
EC-Konformit?tserkl?rung  EC-Declaration of Conformity
EC- Déclaration de conformité  EC-Declaración de Conformidad
EC-Dichiarazione di conformità  EC-Conformiteitverklaring
EC- Declara??o de conformidade  EC- Prohlá?ení o shode
EC-Deklaracja zgodno?ci  EC-Заявление о соответствии
D
Konformit?ts-
erkl?rung
Wir erkl?ren hiermit, dass das Produkt, auf das sich diese Erkl?rung bezieht,
mit den nachstehenden Normen übereinstimmt.
GB
Declaration of
conformity
We hereby declare that the product to which this declaration refers conforms
with the following standards.
CZ
Prohlá?ení o
shode
Tímto prohla?ujeme, ?e vyrobek, kterého se toto prohlá?ení tyká, je v souladu
s ní?e uvedenymi normami.
E
Declaración de
conformidad
Manifestamos en la presente que el producto al que se refiere esta
declaración está de acuerdo con las normas siguientes
F
Déclaration de
conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la
présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di
conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferi-
sce è conforme alle norme di seguito citate.
NL
Conformiteit-
verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking
heeft, met de hierna vermelde normen overeenstemt.
P
Declara??o de
conformidade
Declaramos por meio da presente que o produto no qual se refere esta
declara??o, corresponde às normas seguintes.
PL
Deklaracja
zgodno?ci
Niniejszym o?wiadczamy, ?e produkt, którego niniejsze o?wiadczenie dotyczy,
jest zgodny z poni?szymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
Electronic Balance: KERN KBJ_NM
EU Directive  Standards 
2004/108/EC  EN 55011: 2009+A1:2010
EN 61000-3-2 : 2006-04+A1: 2009 + A2 : 2009
EN 61000-3-3 : 2008
EN 55022: 2010
OIML R 76-1 : 2006
EN45501 : 1992-10+AC : 1993-08
EN 55024 : 2010
(Limit class B)
(Limit class A)
(Limit class B)
2006/95/EC  EN60950 
Signatur
Signature
Datum
Date
19.07.2013
Ort der Ausslung
Place of issue
72336 Balingen  Albert Sauter
KERN & Sohn GmbH
Gesch?ftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, . +49-[0]7433/9933-0
+49-[0]7433/9933-149, : info@kern-sohn.com, Internet: www.kern-sohn。。com
572/573/KB/DS/FKB/FCB/KBJ-BA-e-1471  42

收藏該商鋪

請(qǐng) 登錄 后再收藏

提示

您的留言已提交成功!我們將在第一時(shí)間回復(fù)您~
二維碼 意見(jiàn)反饋
在線留言
临澧县| 山西省| 湘西| 诏安县| 湘乡市| 八宿县| 应用必备| 大安市| 太和县| 资溪县| 罗田县| 襄樊市| 桑植县| 方城县| 蒙城县| 凤阳县| 融水| 济南市| 朝阳区| 保靖县| 霍林郭勒市| 宝山区| 江孜县| 鹤壁市| 花莲市| 宜城市| 崇州市| 清水县| 新河县| 和龙市| 龙州县| 博客| 西乌珠穆沁旗| 井研县| 罗平县| 屏东市| 太白县| 同心县| 丰原市| 墨竹工卡县| 于田县|