Human Corticotropin Releasing
Hormone (CRH)
ELISA Kit
Catalog No. CSB-E06872h
(96T)
This immunoassay kit allows for the in vitro quantitative determination of human
CRH concentrations in serum and other biological fluids.
Expiration date six months from the date of manufacture
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
2
INTRODUCTION
Corticotropin-releasing hormone (CRH) is a 41-amino acid
peptide derived from a 191-amino acid preprohormone. CRH is
secreted by the paraventricular nucleus (PVN) of the
hypothalamus in response to stress. Marked reduction in CRH
has been observed in association with Alzheimer's disease, and
autosomal recessive hypothalamic corticotropin deficiency has
multiple and potentially-fatal metabolic consequences including
hypoglycemia and hepatitis. In addition to being produced in the
hypothalamus, CRH is also synthesized in peripheral tissues,
such as T lymphocytes, and is highly expressed in the placenta.
In the placenta, CRH is a marker that determines the length of
gestation and the timing of parturition and delivery. A rapid
increase in circulating levels of CRH occurs at the onset of
parturition, suggesting that, in addition to its metabolic functions,
CRH may act as a trigger for parturition.
PRINCIPLE OF THE ASSAY
The microtiter plate provided in this kit has been pre-coated with
3
goat-anti-rabbit antibody. Standards or samples are then added
to the appropriate microtiter plate wells with a Horseradish
Peroxidase (HRP)-conjugated CRH and antibody preparation
specific for CRH, and incubated. Then substrate solutions are
added to each well. The enzyme-substrate reaction is terminated
by the addition of a sulphuric acid solution and the color change
is measured spectrophotometrically at a wavelength of 450 nm ±
2 nm. The concentration of CRH in the samples is then
determined by comparing the O.D. of the samples to the
standard curve.
DETECTION RANGE
0.8 ng/ml-40 ng/ml. The standard curve concentrations used for
the ELISA’s were 40 ng/ml, 20 ng/ml, 8 ng/ml, 2.4 ng/ml, 0.8
ng/ml.
SPECIFICITY
This assay recognizes human CRH. No significant
cross-reactivity or interference was observed.
4
SENSITIVITY
The minimum detectable dose of human CRH is typically less
than 1 ng/ml.
The sensitivity of this assay, or Lower Limit of Detection (LLD)
was defined as the lowest protein concentration that could be
differentiated from zero.
MATERIALS PROVIDED
Reagent Quantity
Assay plate 1
Standard 5 x 0.5 ml
HRP-Conjugate 1 x 6ml
Antibody 1 x 6 ml
Wash Buffer
1 x 15 ml
(20×concentrate)
Substrate A 1 x 7 ml
Substrate B 1 x 7 ml
Stop Solution 1 x 7 ml
Standard S1 S2 S3 S4 S5
Concentration
(ng/ml)
0.8 2.4 8 20 40
5
STORAGE
1. Unopened test kits should be stored at 2-8°C upon receipt
and the microtiter plate should be kept in a sealed bag. The
test kit may be used throughout the expiration date of the kit,
provided it is stored as prescribed above. Refer to the
package label for the expiration date.
2. Opened test plate should be stored at 2-8°C in the aluminum
foil bag with desiccants to minimize exposure to damp air. The
kits will remain stable until the expiring date shown, provided it
is stored as prescribed above.
3. A microtiter plate reader with a bandwidth of 10 nm or less
and an optical density range of 0-3 OD or greater at 450nm
wavelength is acceptable for use in absorbance
measurement.
REAGENT PREPARATION
1. Bring all reagents to room temperature before use.
2. Wash Buffer If crystals have formed in the concentrate,
warm up to room temperature and mix gently until the
6
crystals have compley dissolved. Dilute 15 ml of Wash
Buffer Concentrate into deionized or distilled water to prepare
300 ml of Wash Buffer.
Precaution: The Stop Solution provided with this kit is an acid solution. Wear
eye, hand, face, and clothing protection when using this material.
OTHER SUPPLIES REQUIRED
Microplate reader capable of measuring absorbance at 450
nm, with the correction wavelength set at 540 nm or 570 nm.
Pipettes and pipette tips.
Deionized or distilled water.
Squirt bottle, manifold dispenser, or automated microplate
washer.
An incubator which can provide stable incubation conditions
up to 37°C±0.5°C.
SAMPLE COLLECTION AND STORAGE
Serum Use a serum separator tube (SST) and allow
samples to clot for 30 minutes before centrifugation for 15
minutes at 1000 g. Remove serum and assay immediay or
7
aliquot and store samples at -20°C. Centrifuge the sample
again after thawing before the assay. Avoid repeated
freeze-thaw cycles.
Note: Grossly hemolyzed samples are not suitable for use in this assay.
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. It is
recommended that all samples, standards, and controls be assayed in duplicate.
All the reagents should be added directly to the liquid level in the well. The
pipette should avoid contacting the inner wall of the well.
1. Set a Blank without any solution. Add 50µl of Standard or
Sample per well.
2. Add 50µl of HRP-Conjugate and 50µl of Antibody to each
well. Not to Blank well!
3. Cover with the adhesive strip. Incubate for 2 hours at 37° C.
4. Aspirate each well and wash, repeating the process three
times for a total of three washes. Wash: Fill each well with
Wash Buffer (200µl) and let it stand for 2 minutes, then
remove the liquid by flicking the plate over a sink. The
remaining drops are removed by patting the plate on a paper
towel. Complete removal of liquid at each step is essential to
good performance.
8
5. Add 50µl of Substrate A and 50µl of Substrate B to each
well. Incubate for 15 minutes at 37°C. Keeping the plate away
from drafts and other temperature fluctuations in the dark.
6. Add 50µl of Stop Solution to each well. If color change does
not appear uniform, gently tap the plate to ensure thorough
mixing.
7. Determine the optical density of each well within 30 minutes,
using a microplate reader set to 450 nm.
CALCULATION OF RESULTS
Using the professional soft "Curve Exert 1.3" to make a standard curve is
recommended, which can be downloaded from our web.
Average the duplicate readings for each standard, control, and
sample and subtract the average zero standard optical density.
Create a standard curve by reducing the data using computer
software capable of generating a four parameter logistic (4-PL)
curve-fit. As an alternative, construct a standard curve by plotting
the mean absorbance for each standard on the y-axis against
the concentration on the x-axis and draw a best fit curve through
the points on the graph. The data may be linearized by plotting
the log of the CRH concentrations versus the log of the O.D. and
9
the best fit line can be determined by regression analysis. This
procedure will produce an adequate but less precise fit of the
data. If samples have been diluted, the concentration read from
the standard curve must be multiplied by the dilution factor.
LIMITATIONS OF THE PROCEDURE
The kit should not be used beyond the expiration date on the
kit label.
Do not mix or substitute reagents with those from other lots or
sources.
If samples generate values higher than the highest standard,
dilute the samples with the appropriate Diluent and repeat the
assay.
Any variation in operator, pipetting technique, washing
technique, incubation time or temperature, and kit age can
cause variation in binding.
This assay is designed to eliminate interference by soluble
receptors, binding proteins, and other factors present in
biological samples. Until all factors have been tested in the
Quantikine Immunoassay, the possibility of interference
cannot be excluded.
10
TECHNICAL HINTS
Centrifuge vials before opening to collect contents.
When mixing or reconstituting protein solutions, always avoid
foaming.
To avoid cross-contamination, change pipette tips between
additions of each standard level, between sample additions,
and between reagent additions. Also, use separate reservoirs
for each reagent.
When using an automated plate washer, adding a 30 second
soak period following the addition of wash buffer, and/or
rotating the plate 180 degrees between wash steps may
improve assay precision.
To ensure accurate results, proper adhesion of plate sealers
during incubation steps is necessary.
Substrate Solution should remain colorless or light blue until
added to the plate. Keep Substrate Solution protected from
light. Substrate Solution should change from colorless or light
blue to gradations of blue.
11
Stop Solution should be added to the plate in the same order
as the Substrate Solution. The color developed in the wells
will turn from blue to yellow upon addition of the Stop Solution.
Wells that are green in color indicate that the Stop Solution
has not mixed thoroughly with the Substrate Solution.
慧嘉生物您實(shí)驗(yàn)身邊的好伙伴
為客戶提供“zui高品質(zhì)的產(chǎn)品”和“zui的服務(wù)”
AssayBiotech CUSABIO Immunoway Santa Abcam Cst jackson Pierce Sigma Amresco Qiagen Cayman abnova millipore invitrogen merk ebioscience prospec
LifeSpan BD 歡迎廣大客戶咨詢,另有大量宣傳海報(bào)和小禮品贈(zèng)送。
:www.biohj.com
:382603320 1284882975
請(qǐng)輸入賬號(hào)
請(qǐng)輸入密碼
請(qǐng)輸驗(yàn)證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購買風(fēng)險(xiǎn),建議您在購買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。