您好, 歡迎來到化工儀器網(wǎng)! 登錄| 免費注冊| 產(chǎn)品展廳| 收藏商鋪|
The objective of this investigation was to evaluate two methods for measuring the maximum sublimation rate that a freeze-dryer will support—the minimum controllable pressure method and the choke point method. Both methods gave equivalent results, but the minimum controllable pressure method is preferred, since it is easier, faster, and less subjective. The ratio of chamber pressure to condenser pressure corresponding to the onset of chokedflflow was considerably higher in this investigation (up to about 20:1) than in previously published reports. This ratio was not affected by the location of the pressure gauge on the condenser; that is, on the foreline of the vacuum pumpversuson the body of the condenser itself. The total water loss due to sublimation as measured by tunable diode laser absorption spectroscopy was consistently within 5% of gravimetrically determined weight loss, regardless of whether the measurement took place during chokedversusnon-choked process conditions.
請輸入賬號
請輸入密碼
請輸驗證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實性、準確性和合法性由相關(guān)企業(yè)負責,化工儀器網(wǎng)對此不承擔任何保證責任。
溫馨提示:為規(guī)避購買風險,建議您在購買產(chǎn)品前務必確認供應商資質(zhì)及產(chǎn)品質(zhì)量。